EasyManua.ls Logo

bioMetric SpinalPak - Button Functions; ALARM ON;OFF BUTTON; LCD Symbol Descriptions and Instructions

bioMetric SpinalPak
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 9
Button Functions
ALARM ON/OFF BUTTON
The Biomet
®
SpinalPak
®
Non-invasive Spine Fusion Stimulator System is activated as soon as a
chargedbatterypackisinserted.ThebuttonlocatedbelowtheLCDdisplayenablesordisablesthe
audiblealarm.Duringanalarmcondition,depressingthebuttonquickly(0.5seconds)willtemporarily
disabletheaudiblealarm.Depressingthebuttonforalongerperiodoftime(3seconds)willtoggle
theaudiblealarmbetweenenabledanddisabled.Patientsshouldbeadvisedtoleavetheaudiblealarm
enabledasfrequentlyaspossibleinordertoassurethefullyprescribedtreatment.Aspeakersymbol
willbeindicatedontheLCDdisplaywhenthealarmisenabled.
Thealarmdefaultstoaudiblealarm.Pressthebuttonbelowthedisplayonthefrontofthestimulator
tosilencethealarm.Aftersilenced,thelightwillcontinuetoflashandthedisplaywillindicatethe
alarmcondition.
Symbol Condition Instructions
Treating Continueuse.
Audiblealarm Ifbeeping,depressthebutton
brieflytosilencethealarm.Depressthe
buttonapproximately3secondsto
engageordisengagetheaudiblealarm.
Lowbatterycharge Insertachargedbatterypack.
Disconnection Confirmthateachelectrodeisproperly
appliedontheskin.Seetheelectrode
pouchforinstructions.Confirmthatthe
leadwireisattachedproperly.Replace
theleadwireifnecessary.
Systemerror Errorinthestimulator–
ContactBiometforassistance.
Stimulatoris StimulatorwillnottreatuntilUSB
connectedto cableisdisconnected
a PC
Endofoperation/ ContactBiomet
TreatmentCompletion
LCD Symbol Descriptions and Instructions