M
button repeatedly to view all the measurements that are recorded on
the device.
3.9AMAveraging(Morning)
This feature averages the morning readings between 4am and 12pm to identify morning hypertension. Several
studies have shown elevated cardiovascular risks, heart attacks, and strokes occur in the morning as the body is
waking up from sleep. See Section 3.8 on how to view the average value of the morning readings.
3.10 Clearing All Values
If you are sure that you want to permanently remove all stored values, hold down the
M
button
when in standby (nothing is displayed on the screen) until the “CL” appears and then release the
button. To permanently clear the memory, press the
M
button while “CL” is flashing.
4. Error Messages / Malfunctions
If an error occurs during a measurement the LCD displays the corresponding error code.
Error Possible Cause Remedy
EE 1 Cuff pressure is more than 300 mmHg or the
inflation of the cuff takes too long.
Ensure that the cuff is worn correctly and mea-
sure again. Avoid movement or talking when
the cuff is being inflated.
EE 2 The difference between systolic and diastolic
pressure readings are out of reasonable range.
Ensure that the cuff is being worn correctly and
that you have been inactive for a sufficient time
before making the measurement.
EE 3 Unnatural pressure impulses influence the mea-
surement result. Reason: the arm was moved
during the measurement.
Avoid unnecessary movement or talking.
Err 5 No pulse has been detected. Ensure the cuff is being worn correctly, and that
you have your arm at the heart level.
EE E Memory record.
Low battery Replace batteries.
If problems occur when using the device the following points should be checked, and if necessary, the
corresponding measures should be taken.