Conditions for MR scans with a BIOTRONIK ProMRI
®
system (continued)
Requirements of the MRI scanner
Use of a clncal MRI machne wth
A closed tube
ylndrcal magnets
A statc magnetc eld strength of 15 tesla
The slew rate of the MRI scanner’s gradent elds should not exceed
200 T/m/s per axs
Restrictions during the MRI scan
The mean specc absorpton rate (SAR) for the whole body dsplayed
by the MRI scanner must not exceed 20 W/kg
The head absorpton rate dsplayed by the MRI scanner must not
exceed 32 W/kg
Emergency equpment for resusctaton must be kept at hand and
properly certed staff must be avalable
ontnuously montor the patent’s condton durng the entre MR
scan usng at least one of the followng parameters blood oxygen
saturaton, blood pressure or E
Note The E functon ntegrated n the MRI scanner s often not
permtted for patent montorng Therefore, only use devces whch
are permtted for patent montorng n an MRI envronment