Checklist before MRI scanning
This checklist will help to ensure the safe application of an MR scan on
patients with a BIOTRONIK device system that has been labeled MR
Conditional.
*
We recommend checking the boxes off while using this checklist to
ensure that an MR Conditional scan with BIOTRONIK ProMRI
®
systems
is permissible. Please check to see whether the patient’s ProMRI
®
device-lead combination is approved for a full body MR scan.
(See charts on p. 10.)
General requirements for patients with ProMRI
®
systems
Use of a clncal MRI system wth a closed tube, cylndrcal
magnets and a statc magnetc eld strength of 15 tesla
The slew rate of the MRI scanner’s gradient fields should not
exceed 200 T/m/s per axis.
The mean specc absorpton rate for the whole body dsplayed by
the MRI scanner must not exceed 20 W/kg
The head absorpton rate dsplayed by the MRI scanner must not
exceed 32 W/kg
Emergency equpment for resusctaton must be kept at hand and
properly certed staff must be avalable