DISASSEMBLY
AND.
REASSEMBLY
INSTRUCTIONS
SECTION
6:
SECTION
G
DISASSEMIBLY
AND
REASSEMBLY
INSTRUCTIONS
The
Bird
respirators
have
been
designed
to
pro-
vide
the
maximum
amount
of
utilization
with
a
minimum
of
maintenance.
These
instructions
will
provide
the
information
necessary
to
disassemble
and
reassemble
the
Bird
MARK
7
or
MARK
7A
Respirator.
WARNING:
THE
BIRD.
MARK
7
ANDIOR
7A
SHOULD
BE
SERVICEDICALIBRATED
BY
A
BIRD
PRODUCTS
CORPORATION.
TRAINED
HOSPITALIDEALER
SERVIGE
TECHNICIAN.
D
To
properly
service
the
Bird
respirators,
a
service
tool
kit,
P/N
7723,
is
required.
The
following
special
tools
are
included
in
P/N
7723
and
are
required
for
calibration
and
overhaul:
PART
#
DESCRIPTION
P/N
0031
Ceramic
Bushing
Tool
P/N
0042
Lubewick
PN
0631
Silicone-Base
Lubricant
P/N
0924
Styiet
P/N
0916
|
Mainstream
Stopper
(22mm)
PIN
5497
Sensitivity
Adjustment
Tool
P/N
6711
Tube
Removal
Tool
P/N
6754
Calibration
Regulator
PIN
6855
Orificed
Adapter
In
addition,
the
following
commercially
available
tools
are
required:
5/6"
Nut
Driver
%s"
Allen
Wrench
362"
Allen
Wrench
%
Open
End
Wrench
78"
Open
End
Wrench
Truarc
Pliers
#1
Thin
Bladed
Screwdriver
¥2"-7he"
Open
End
Wrench
Va
58"
Open
End
Wrench
Diagonal
Cutting
Pliers
Needlenose
Pliers
4.
DISASSEMBLY
1.0
AMBIENT
COMPARTMENT
REMOVAL
1.1
Remove
(2)
socket
screws
and
washers
using
%e”
allen
wrench
and
carefully
pull
Ambient
Compartment
off
until
the
hand
timing
rod
clears
the
magnet
(Fig.
22).
NOTE:
To
facilitate
removal
of
Ambient
Compartment,
it
will
be
necessary
to
dislodge
the
venturi
tube
assembly
from
the
ambient
filter
inlet
port.
FIGURE
22
1.2
Using
tube
removal
tool
(P/N
6711),
separate
manometer
tube
#25
from
quick
disconnect
at
the
Centerbody.
1.3
To
remove
manometer
from
Ambient
Com-
pariment,
use
%8"
open
end
wrench
to
allow
separation
of
mounting
bracket
from
manome-
ter,
then
remove
manometer.
1.4
Using
а
%2"
allen
wrench,
loosen
the
two
(2)
setscrews
securing
the
Sensitivity
Radius
Arm
to
the
magnet,
then
remove
the
arm.
1.5
Remove
the
magnet
from
the
Ambient
Com-
partment
by
rotating
the
assembly
in
a
COUNTERCLOCKWISE
direction.
1.6
Disassemble
the
Ambient
Filter
as
follows:
=
Using
a
Phillips
screwdriver,
remove
the
screw
securing
the
filter
holder
to
the
filter
housing.
m
Using
your
fingers
ONLY
separate
the
remaining
components
for
proper
cleaning.
2.0
PRESSURE
COMPARTMENT
REMOVAL
2.1
Use
546”
nut
driver
to
remove
inspiratory
power
drive
line
stud
and
nut
seal.