E14C4t User Guide (16-01265-EN-07) Reprocessing 7
Service and Repair
Caring for the Transducer
The transducer may be damaged during use or reprocessing, so it must be checked
before use for cracks or irregularities in the surface, following the procedure in Care
and Cleaning. It should also be checked thoroughly once a month following the
same procedure.
Reprocessing
To ensure the best results when using BK Medical equipment, it is important to
maintain a strict cleaning routine.
Complete details and procedures can be found in Care and Cleaning that
accompanies this user guide. A list of reprocessing methods that the transducer can
withstand are listed in the Product Data Sheet.
Sterile covers are available. See the Product Data Sheet for more details.
Starting Imaging
Before use, all equipment must be reprocessed according to expected use.
WARNING GS-w5
Do not run over transducer cables with system wheels or other heavy equipment. This can
damage the cable, and lead to degradation and/or corruption of the ultrasound image.
WARNING AO-w1
To avoid tissue damage, always keep the exposure level (the acoustic output level and the
exposure time) as low as possible.
WARNING SR-w1
Service and repair of BK electromedical equipment must be carried out only by the
manufacturer or its authorized representatives. BK Medical reserves the right to disclaim all
responsibility, including but not limited to responsibility for the operating safety, reliability
and performance of equipment serviced or repaired by other parties. After service or repairs
have been carried out, a qualified electrician or hospital technician should verify the safety of
all equipment.
WARNING Reproc-w2
Users of this equipment have an obligation and responsibility to provide the highest possible
degree of infection control to patients, co-workers and themselves. The instructions in this
book are meant as a guide. To avoid cross-contamination, follow all infection control policies
(including for reprocessing, packing and storage) for personnel and equipment that have
been established for your office, department or hospital.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Medical E14C4t and is the answer not in the manual?

BK Medical E14C4t Specifications

General IconGeneral
BrandBK Medical
ModelE14C4t
CategoryMedical Equipment
LanguageEnglish

Related product manuals