Why is my bluelab pH Pen Measuring Instruments calibration unsuccessful?
- VVanessa FisherAug 21, 2025
An unsuccessful calibration of your Bluelab Measuring Instruments may occur due to several reasons: * The buffers you are using might be inaccurate, in this case, replace them. * The glassware might not be clean, ensure you clean the glassware thoroughly. * The glassware could be aged; if cleaning doesn't help, replace the unit. * The probe might not be hydrated. Soak the probe in KCl storage solution for 24 hours, then retest.