EasyManua.ls Logo

Bluewater Pro 400C-HF - Technical Information

Bluewater Pro 400C-HF
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
07
© Bluewater Group 2016
EN
Technical information
Size W×H×D Weight Power Power supply
earthed
Purified water flow
rate
Sound level
Pro 400B-HR
230V
225×466×460 mm 29 kg 530 W 230V, 50-60 Hz 3.1 ± 0.5 l/min
< 60 dB(A)
120V
8.9×18.3×18.1 “ 64 lbs 6 Amp 120V, 60 Hz 0.8 ± 0.13 gal/min
Pro 400C-HF
230V
225×466×460 mm 29 kg 530 W 230V, 50-60 Hz 3.5 ± 0.5 l/min
< 60 dB(A)
120V
8.9×18.3×18.1 “ 64 lbs 6 Amp 120V, 60 Hz 0.9 ± 0.13 gal/min
Pro 400C-HR
230V
225×466×460 mm 29 kg 530 W 230V, 50-60 Hz 3.1 ± 0.5 l/min
< 60 dB(A)
120V
8.9×18.3×18.1 “ 64 lbs 6 Amp 120V, 60 Hz 0.8 ± 0.13 gal/min
Pro 400BC-HR
230V
225×466×460 mm 29 kg 530 W 230V, 50-60 Hz 3.1 ± 0.5 l/min
< 60 dB(A)
120V
8.9×18.3×18.1 “ 64 lbs 6 Amp 120V, 60 Hz 0.8 ± 0.13 gal/min
Pro 400CV-HR
230V
225×466×470 mm 29 kg 530 W 230V, 50-60 Hz 3.0 ± 0.5 l/min
< 60 dB(A)
120V
8.9×18.3×18.5 “ 64 lbs 6 Amp 120V, 60 Hz 0.8 ± 0.13 gal/min
Pro 400BCV-HR
230V
225×466×470 mm 29 kg 530 W 230V, 50-60 Hz 3.0 ± 0.5 l/min
< 60 dB(A)
120V
8.9×18.3×18.5 “ 64 lbs 6 Amp 120V, 60 Hz 0.8 ± 0.13 gal/min
Pro 600C-HF
230V
225×466×460 mm 29 kg 720 W 230V 50-60 Hz 5.2 ± 0.5 l/min < 60 dB(A)
Pro 600CV-HR
230V
225×466×470 mm 29 kg 720 W 230V 50-60 Hz 4.5 ± 0.5 l/min < 60 dB(A)
Requirements on feed water to water purifier
Temperature
Feed water flowing
pressure
Water flow
TDS
(Total Disolved Solids)
[ppm]
Conductivity
[µS/cm, 25 °C
(77 °F)]
Hardness
Iron Fe
(+2)
/Fe
(+3)
[mg/l]
Turbidity [FNU]
Pro 400B-HR
Pro 400BC-HR
Pro 400BCV-HR
230V
2-30 °C 1.5-10 bar > 8 l/min
< 7500 < 13 000
< 20 °dH
< 1.5/< 0.3 < 0.5
120V
36-86 °F 22-145 psi > 2.1 gal/min < 20.9 grains/gal
Pro 400C-HF
Pro 400C-HR
Pro 400CV-HR
230V
2-35 °C 1.5-10 bar > 8 l/min
< 1500 < 2000
< 20 °dH
< 1.5/< 0.3 < 0.5
120V
36-95 °F 22-145 psi > 2.1 gal/min < 20.9 grains/gal
Pro 600C-HF
Pro 600CV-HR
230V
2-35 °C 2-10 bar > 10 l/min < 1500 < 2000 < 20 °dH < 1.0/< 0.3 < 0.5
Do not install the water purifier if the feed water is microbiologically unsafe!
For more detailed information see www.bluewatergroup.com