Auto CPAP / CPAP System User Manual V1.0
34 / 51
16.2 Options of the Patient Menu and Corresponding
Descriptions
There are five humidity levels available. As the
numbering increases, the humidity rises accordingly.
―Off‖ means the humidifier is turned off.
Set humidifier to preheat by adjusting this option.
This feature automatically turns off after 30 minutes
There are five heat levels available. As the
numbering increases, the heat rises accordingly.
―Off‖ means the heat is turned off. Note: Heated
Tubing is displayed in the patient menu only when it
is connected
In order to increase comfort and help the patient fall
asleep easily, the pressure can increase gradually,
when the Ramp feature is enabled. The ramp time
during which the initial pressure rises to the
prescribed treatment pressure can be adjusted. As
you turn the Knob to the nearest point, the
numbering increases or decreases by five seconds.
The screen displays a real-time countdown of the
remaining ramp time in seconds
When the humidifier is on, this feature allows the
airflow to continue for about 15 minutes at a low
pressure (about 2 hPa) after you press the Start /
Stop to discontinue treatment. This will blow
off the vapor left in the water chamber to avoid any
damage to the device. When this feature is set to
―Off,‖ which means it is disabled, the airfolw stops
delivering air instantly after you press the Start /
Stop .
Set date by adjusting this option
yy mm dd /
mm dd yy /
dd mm yy
Turn the Knob to choose among three date
formats.
Set time by adjusting this option
Turn the Knob to choose between the two
time formats.
Setting screen brightness by adjusting this option.‖