M1 Mini User Manual
5 / 32
Static Pressure Stability
± 0.5 hPa
Ramp
The ramp time ranges from 0 to 60 minutes.
Sound Pressure Level
< 30 dB, when the device is working at the pressure of 10 hPa.
Sound Power Level
< 38 dB, when the device is working at the pressure of 10 hPa.
Maximum Flow
Test Pressures (hPa)
4
8
12
16
20
Measured Pressure at the
Patient Connection Port (hPa)
3
7
11
15
19
Average Flow at the Patient
Connection Port (L/min)
85
125
110
110
95
When the working pressure is set to the values listed in the table, the average
flow rate at the patient end should be greater than 80% of the corresponding
flow value in the table.
Air Tubing
Air tubing
Length
Inner diameter
Tubing
6 ft.(1.83 m)
15 mm
Tubing
6 ft.(1.83 m)
19 mm
The Form and the Dimensions of the Patient Connection Port
The 22 mm conical air outlet complies with ISO 5356-1.
Air filter
Filtration efficiency: > 20% for 10 micron
Material: Non-woven fabric and Polyester
Aircraft use
BMC confirms that the device meets the Federal Aviation Administration (FAA) requirements
(RTCA/DO-160, section 20, category T and section 21, category M) for all phases of air travel.
Bluetooth authentication information
Product name: Auto CPAP System
Model: M1 Mini
DID: D051679
QDID: 154506