M1 Mini User Manual
27 / 32
25. EMC Requirements
The cables must be provided by BMC. The information of each cable is as follows:
(1) Power adapter: 1800 mm ± 45 mm, unshielded
(2) Tubing: 1800 mm ± 10%, unshielded
Guidance and manufacturer's declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should ensure that it is used in such an environment.
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for
power supply lines
± 2 kV for
power supply lines
± 1 kV line (s) to line (s)
± 1 kV line (s) to line (s)
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0%
U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0%
U
T
; 1 cycle
70 %
U
T
; 25/30 cycle;
At 0°
0%
U
T
; 250 / 300 cycle
0%
U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0%
U
T
; 1 cycle
70 %
U
T
; 25/30 cycles
At 0°
0%
U
T
; 250 / 300 cycle
Power frequency (50
/ 60 Hz) magnetic
field
IEC 61000-4-8
NOTE:
U
T
is the a.c. mians voltage prior to application of the test level.