Service Manual for RESmart
®
Auto CPAP V1.1
10/16
Components Illumination
1. Device Inside
S/N Assembly Figure
1 Insert humidifier locker on the shield bottom.
2
Insert humidifier power supply reed on the
shield bottom as per Fig. E and F.
3
Assembly the humidifier locker and button
spring.
4
Put fixing on the spring and fix by screw (2×
3mm*8mm).
5 Put power wire.
6
Assembly the shield back on the bottom, fix by
screw (2×3mm*8mm).
7 Assembly foam (bottom) into shield bottom.
8
Assembly AC power socket on the shield back
(character face up).
9 Clean up the wire.
10
Assembly power PCBA in the socket and
connect all plugs. Clean up the wire.
11 Connect outlet to blower connector, fix by clip.