EasyManua.ls Logo

BMC RESmart Auto CPAP - Appendix B: Spare Part Configuration

BMC RESmart Auto CPAP
19 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual for RESmart
®
Auto CPAP V1.1
13/16
2
Put spring on each of three poles on heater
plane.
3
Assembly the scaleboard on the heater plane
and fix by screw (3mm*8mm).
Scaleboard
4
As per Table 3, assembly heater PCBA and
wires.
5
Put the humidifier inner connector on the
shield.
6
Put on light window.
7
Assembly the inner and outer shield.
8
Fix on screw (M3*10mm).
Appendix B:
Spare Part Configuration
S/N Name Qty Figure
1 Shield Cover 1
2 Handle 1
3 Handle Fixing 2
4 Shield Bottom 1

Related product manuals