Service Manual for RESmart
®
Auto CPAP V1.1
12/16
2. Device Outside
S/N
Assembly Figure
18
Put handle in the socket on shield cover.
19
Assembly handle fixing to handle and fix by
screw (4×3mm*8mm).
20
Assembly the infrared window.
21
Assembly and fix the panel.
23
Put on the key button.
24
Assembly shield cover on the bottom, adjust
the position of outlet and make sure it is
clamped in the right socket.
25
Fix the shield by screws (4×M3*10mm).
26
Stick four device pads.
27
Put in filter and assembly the cover.
28
Put on all plugs if necessary (without
humidifier).
3. Humidifier
S/N
Assembly Figure
1
As per Fig. H, assembly the heater plane and
wire. Fix them by screw (3mm*8mm).