RESmart GII BPAP System Y / U Series User Manual V2.1
38 / 42
Guidance and manufacturer's declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
user of the device should make sure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge
(ESD)
IEC
61000-4-2
± 8 kV contact
± 15 kV air
± 68 kV contact
± 15 kV air
Floor should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%
Electrical fast
transient /
burst
IEC
61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input / output
lines
± 2 kV for
power supply
lines
± 1 kV for
input / output
lines
Mains power quality should be
that of a typical commercial or
hospital environment
Surge
IEC
61000-4-5
± 1 kV
line (s) to line (s)
± 2 kV
common mode
± 1 kV
line (s) to line (s)
± 2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
0%
U
T
; 0.5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0%
U
T
; 1 cycle
70%
U
T
; 25 / 30
cycle
At 0°
0%
U
T
; 250 /
300 cycle
0%
U
T
; 0.5
cycle
At 0°, 45°, 90°,
135°, 180°,
225°, 270° and
315°
0%
U
T
; 1 cycle
70%
U
T
; 25 / 30
cycle
At 0°
0%
U
T
; 250 /
300 cycle
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the device requires continued
operation during power mains
interruptions, it is recommended
that the device be powered from
an uninterruptible power supply
or a battery
Power
frequency (50
/ 60 Hz)
magnetic field
IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Note:
U
T
is the a.c. mains voltage prior to application of the test level.