H60 Heated Humidifier User Manual V2.0
13 / 21
Fig. 5-11
(4) Connect the top cover and main body of the humidifier properly, as shown in the figure
below.
Fig. 5-12
6. Disinfection
Generally speaking, if you have strictly followed the above cleaning instructions, you do not
have to disinfect the humidifier. If the device is contaminated or used in clinical trials, you
may purchase disinfectants from a pharmacist to disinfect the water chamber.
Disinfection of Humidifier Water Chamber
Prior to disinfection, clean the water chamber according to Section 5.4 “Cleaning the Water
Chamber”. The disinfection methods are as follows:
(1) Heat disinfection: Disinfect the water chamber by immersing it in tap water at 75°C±2°C
for 30 minutes.
(2) Use mild disinfectants.
CAUTIONS!
• Disinfectants tend to damage materials and reduce the life of components. Try to select the
appropriate disinfectant, and follow the disinfectant manufacturer's instructions and
recommendations.
• After disinfection, check the disinfected component for any signs of damage. Replace any
damaged component immediately.
WARNINGS!
• After disinfection, rinse any disinfected component in clean water thoroughly, especially