OZONE MONITOR BMT 932 Manual, Rev. 02/2021
24
This is an example for changing a logical variable: the unit of ozone concentration shall be
changed from ppm to µg/m³:
Button Display
0.000 ppm
SCROLL
View Parameters
SCROLL (>2 s)
Set Parameters
ENTER
Set Ozone Unit
ENTER
Ozone: ppm
SET
Ozone: µg/m³
ENTER
Ozone: µg/m³
Saving...
BACK
Set Ozone Unit
BACK
Set Parameters
BACK
0 µg/m³
One more example: the numerical parameter High Alarm - Threshold (determines the thresh-
old, above which a High Alarm occurs) shall be changed from 0.3 ppm to 0.52 ppm only for
channel 2:
Button Display
0.000 ppm
SCROLL
View Parameters
SCROLL (> 2s)
Set Parameters
ENTER
Set Ozone Unit
SCROLL
Set Alarms
ENTER
Set All High Alarms
SCROLL
Set High Alarm Ch1
SCROLL
Set High Alarm Ch2
ENTER
Enable/Disable
SCROLL
Set Hi-Threshold
ENTER
Hi:0.300 ppm
ENTER
Hi:0.300 ppm
SET
Hi:0.400 ppm
SET
Hi:0.500 ppm
ENTER
Hi:0.500 ppm
SET
Hi:0.510 ppm
SET
Hi:0.520 ppm
ENTER
Hi:0.520 ppm