EasyManua.ls Logo

Boeing 747 - Page 33

Boeing 747
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
I. Adjust the instrument gain controls to get the notch signal at 80% full screen height. Turn the
transducer in the guide hole to make sure the notch signal is at a maximum amplitude and adjust the
setscrew so that it is tight. Move the transducer guide above the reference notch one more time to
make sure the signal is at a maximum height of 80%. See Figure 7 or Figure 8 as applicable.
J. Put the transducer at the position shown in Table 1 for the back surface signal. Make sure the back
surface signal is at the position shown in Table 1. See Figure 7 or Figure 8 as applicable.
K. Increase the gain 6 dB.
L. Examine all noise signals to make sure the signal-to-noise ratio is larger than 3:1. If the signal-to-noise
ratio is less than 3:1, try a different transducer or a higher frequency transducer.
NOTE: Make sure the initial pulse is not greater than 10% of full screen width. If necessary, adjust the
instrument controls or use a different transducer.
5. Inspection Procedure
NOTE: Each fuse pin must be examined from the inboard and outboard ends of the pin. See Figure 1.
A. Prepare for the inspection as specified in Paragraph 3.
B. Do the instrument calibration as specified in Paragraph 4.
C. Put couplant on the inspection area of the fuse pin. See Figure 9.
D. Put the transducer guide on the fuse pin and make sure the signal from the back surface agrees with
Table 2 and Figure 9. If you do not get a back surface signal, move the transducer to a different
position and try again.
NOTE: If you do not get a back surface signal, check to make sure there is good transducer contact. If
the transducer has good contact, add more gain to get a signal from the back surface that is
approximately 100% full screen height.
E. Make a scan of the fuse pin. As you make a scan, monitor the ultrasonic signals which occur between
10% and 30% full screen width. Also monitor the first signal from the back surface to make sure you
keep full transducer contact.
NOTE: A loss of the back surface signal can be caused by a cracked fuse pin.
F. Do Paragraph 5.C. thru Paragraph 5.E. again on the opposite end of the fuse pin.
G. Refer to Paragraph 6. to make an analysis of the ultrasonic signals that occurred during this inspection.
6. Inspection Results
NOTE: (1) Some fuse pins can have an ultrasonic signal that is out of the area of interest but stays on the
screen during the full inspection of the fuse pin. This signal can occur because of mode conversion
of the ultrasonic sound waves. If the signal is caused by mode conversion, the signal will be small in
amplitude while the ultrasonic signal from the back surface will be large. A crack will usually cause
the ultrasonic signal from the back surface to decrease below 100% full screen height and a signal
from the crack will occur at a lesser screen range.
(2) It is possible that an off angle crack will not give a return echo, but will cause a loss of signal
from the back surface.
A. A fuse pin that does not have a crack indication is acceptable.
B. Do more analysis on a fuse pin with an ultrasonic signal that is 40% or more of full screen height and is
between 10 and 30% full screen width. Refer to Table 2 for the expected crack locations. Do more
analysis as follows:
747
NONDESTRUCTIVE TEST MANUAL
PART 4 54-30-06
Page 3
Nov 15/2015D6-7170
ECCN 9E991 BOEING PROPRIETARY - Copyright © Unpublished Work - See title page for details
EFFECTIVITY
ALL

Other manuals for Boeing 747

Related product manuals