39 Operating Your Ultrasonic Bath
3 Hand tighten the drain valve onto the drain pipe over the white teflon sealing
tape. Finish tightening the valve in place using an adjustable or a 21mm
wrench. Tighten the valve no more than one full turn when using the wrench
until the handle is on top.
CAUTION: Over tightening of the valve can cause damage to the ultrasonic
tank. Always use teflon sealing tape or a sealing paste designed for use with
stainless steel if retightening or refitting of the drain valve is required.
4 Hand tighten the hose adaptor into the end of the drain valve. Slide the drain
tube over the barbed hose adaptor end.
5 Close the drain valve by turning the handle perpendicular to the valve body and
the unit is ready to fill with solution. To open the valve and drain the tank, turn
the handle so that it is in line with the valve body.
Step Action
Protective Cap
Hose
Adaptor
Drain Tube
Valve
Assembled Drain