Displayed temperature range: 93.2 – 108 °F (34 – 42.2 °C)
Operating ambient temperature range: 50 – 104 °F (10 – 40 °C)
Display resolution: 0.1 °F or °C
Accuracy for displayed temperature
range
Maximum Laboratory Error
95 °F – 107.6 °F (35 °C – 42 °C): ± 0.4 °F ± 0.2 °C
outside this range: ± 0.5 °F ± 0.3 °C
clinical repeatability: ± 0.26 °F ± 0.14 °C
Long term storage ranges
Storage/transport temperature: -13 °F to 131 °F (-25 °C to 55 °C)
Humidity: 15-95% non condensing
Battery life: 2 years / 350 measurements
Service life: 5 years
This thermometer is specified to operate at 1 atmospheric pressure (700-1060hPA) or at altitudes with an atmospheric
pressure up to 1 atmospheric pressure (700-1060hPA).
Equipment with type BF applied parts
See instruction for use
Operating temperature
Keep it dry
Attention, consult ACCOMPANYING DOCUMENTS

Other manuals for Braun IRT 6020