The ME equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME equipment should ensure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Radiated RF IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside the shielded
location from fixed RF transmitters, as
determined by an electromagnetic site
survey, should be less than 3 V/m.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Separation distance calculation provided
above. If a known transmitter is present
the specific distance can be calculated
using the equations.
Conducted RF IEC 61000-4-
6
3Vrms 150kHz to
80MHz
Not Applicable (no
electrical cabling)
Electrical fast transient IEC
61000-4-4
±2kV power line ±1kV
I/O lines
Not Applicable
The ME equipment is solely battery
powered.
Surge IEC 61000-4-5
±1kV differential
±2kV common
Not Applicable
Power frequency magnetic
field IEC 61000-4-8
3 A/m Complies
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.