31
Useful information while pump is running
The line describes the course of Cp over the time and the area describes the course of
Ce over the time. The time window is 20 min (15 min past, 5 min future).
Additional information is left with
C.
Finishing TCI
There are two possibilities to finish the TCI Therapy (reversion of anaesthesia or
sedation):
Set Target= 0
Stop pump
It is recommended to simply stop the pump by pressing
sf.
Pressing
R the pump offers additional information in this case the
information is modified the following way:
After the therapy is ended there are two possibilities:
a) The pump may be used for TCI with the same drug again but with a new patient.
In this case, cancel old therapy and use new disposables.
b) The pump may go with the patient but in continuous mode (without TCI).
By pressing r additional
information can be requested.
Pressing r a second time is offering
a graphical overview.
Pressing r again shows up the
graph.
In case a) press u in the menu - in
case b) press
d.
In both cases the “old” TCI needs to
be ended by c and selecting “Yes”
in this screen by pressing u.
Chapter 3
SPECIAL FUNCTIONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml