30
Important notes:
•
Be sure to enter the data corresponding to the respective patient.
•
Once the TCI is started patient data can not be altered!
Editing a target and starting TCI
The editor window for setting the target comes up with the default value from the
drug list.
Confirm target with
k. TCI can be started now with sf.
After TCI is started the screen looks the following:
In the top line there is an icon indicating the parameter set and the mode (Mode
Indicator) with following meaning:
•
“TCI Ma P”: TCI Marsh plasma targeting
•
“TCI Sc P”: TCI Schnider plasma targeting
•
“TCI Sc E”: TCI Schnider effect-site targeting
•
“TCI Mi P”: TCI Minto plasma targeting
•
“TCI Mi E”: TCI Minto effect-site targeting
In the bottom line the status parameters like flow rate, Cp/Ce, infused volume etc. can
be displayed. The desired parameter can be selected by using
t . It is recommended to
select Cp/Ce.
In case it is necessary to change the target press
l to edit the value.
Editing this parameter is guided by
the dose error reduction system
“DoseGuard™” according to the limits
defined in the drug list.
Drug short name
Target and unitCurrent plasma an effect-site concentration
Mode-Indicator
Chapter 3
SPECIAL FUNCTIONS