Chapter 3
SPECIAL FUNCTIONS
Important note: Before installing an additional drug list please contact your local
B. Braun representative!
Setting up the pump
For TCI a drug list with at least one drug activating the profile TCI is necessary. The
drug list in this version is pre-defined. By this the conditions for an effective and safe
therapy are defined.
Switch on pump with
o and wait until self-check is finished. Insert disposable and
use the drug lib according to Instructions for Use.
Selecting a drug
Select drug list, category (the TCI drugs need to be selected from the category “TCI”)
and desired drug by using
q .
These steps are only necessary in case there are different options for that drug.
Input of patient data
Depending on the parameter set one or more of the following data are necessary:
Weight
Height
Gender
Age
The editing window appears with the initial setting “0to make sure editing a value ta-
kes place (exemption: initial setting for gender is “male”).
Using effect-site targeting the weight may be limited due to the constraints of the
LBM calculation.
In this example: Propofol.
As a next step select the correct
dilution (concentration) of the drug
to be administered as well as the
parameter set (model) and the Mode
(Effect-Site Targeting or Plasma
Targeting)
Use q for editing the patient data.
Example.