EasyManuals Logo

Braun Perfusor Space User Manual

Braun Perfusor Space
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
38
Starting Intermittent Mode via Special Function Menu:
Switch on pump with
sf and wait until self-check is finished.
Insert disposable.
Go to Special Functions Menu and select Intermittent Mode.
Press
l to enter parameters and k to confirm.
After entering all desired parameters the pump can be started by pressing
sf
In the upper part of the display the icon for “Intermittent Mode” appears.
In bolus phase the screen shows the following:
After the bolus phase the pump switches to rate phase and the screen shows
the following:
Note: To cancel bolus infusion in the Intermittet Bolus therapy at any time it is
only possible with
sf.
Note: The delivery of drugs can be stopped and started again in the Intermittent
Mode at any time by pressing
sf. During infusion it is possible to change the
bolus volume, amount, VTBI as well as the time interval.
Press
C.
Use
t to navigate through the parameter list and select the parameter to
be changed with
l.
Enter the new value and press
k.
The pump continues infusion.
Changing the bolus after start:
If the user edits the bolus the therapy progression changes.
Press
C.
Use
t to select Bolus and press l.
Change Bolus by using
t and press k to confirm.
The pump automatically recalculates all other settings of the therapy.
The pump now delivers the
predefined bolus.
The pump now delivers the
predefined rate.
SPECIAL FUNCTIONS
Chapter 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml