52
Note:
The OccluGuard status can be checked in the status menu
Perfusor® Space continues to infuse during OccluGuard and Pressure
leap/drop alarms.
The existing occlusion alarm pressure levels are unaffected by OccluGuard.
OccluGuard
Symbol
Meaning
OccluGuard is active.
Infusion is running
stably
Recommendation
n/a
Pending OccluGuard
has not enough data
n/a
OccluGuard Inactive
OccluGuard will automatically
reactivate as soon as infusion rate
drops below threshold levels see
above.
Occlusion has been
detected
Confirm alarm and check IV access, IV
setup and syringe for cause of
occlusion. Should the cause of the
alarm be removed, the alarm will stop
automatically.
OccluGuard is
deactivated
Activate OccluGuard see below(no symbol)
Pressure rise
detect
Pressure
leap/drop
Symbol
Meaning
Sudden pressure leap
detected
Recommendation
Check IV access and IV setup
Sudden pressure drop
detected
Check IV access and IV setup
Pressure leap
detect.
Pressure drop
detect.
Chapter 5
OPTIONS

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml