Chapter 6
ALARMS
Calibrate device Pump calibration data have changed (e.g. after an
update).
Recalibrate device via the service program.
Claw malfunction“ The emergency release button was pressed and the
claws manually opened.
Take out syringe and contact technical service department.
Plunger plate not prop. fixed“ The syringe plunger plate does not attach to the
plunger plate sensor on the pump.
Check system for negative pressure and eliminate
cause. Consider “Patient Safety“.
Standby Time expired“ The set standby time has ended.
Set new time or continue with previously set therapy.
No battery inserted“ It is not possible to use the pump without a
battery pack.
Turn off pump and insert battery pack according
to describtionOverview Perfuso Space.
Data were reset” Therapy and pump settings could not be restored.
Enter therapy and pump settings anew.
"Therapy data were reset" Therapy data could not be restored.
Enter therapy anew.
Data Lock" An attempt was made to stop or switch the pump off
without entering the code.
Enter the correct code in order to continue the
therapy or in order to turn the pump off.
Connection lost TOM aborted Data connection between TOM pumps in the
SpaceStation has been lost and TOM has been aborted
(Take Over Mode only).
TOM may be reactivated by pressing
k and then u
when prompted Return to Take Over M.
Infusion taken over by
other pump Infusion been handed over to second Perfusor® Space
pump (Take Over Mode only)
The red LED extinguishes with the acknowledgement of the alarm.
Caution: If a wrench is displayed and/or a yellow, red and blue LED blink then the
pump is in the service mode and is not permitted to be used on a patient. The pump is
then to be checked by a service technician.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml