ALARMS
the alarm message, and the staff call (optional) are all cleared by pressing k.
Corrections should be made in accordance with the alarm reason.
Display message Alarm reason
Syringe empty“ There is no fluid left in the syringe.
Due to varying syringe tolerances of syringes
from other manufacturers, some fluid may be
left inside the syringe.
Restarting the infusion leads to a complete depletion
of the syringe and shut-off via the pressure sensor.
Perform syringe change as described in 1.4.
VTBI infused The preselected volume was infused.
Continue therapy or select new therapy.
Time expired“ The preselected time has ended.
Continue therapy or select new therapy.
Battery empty“ The battery pack is discharged.
Connect device with mains and/or exchange
battery pack. The battery alarm will be on for 3 min.
Then the pump will automatically turn off.
KVO finished” KVO is reached. Continue with old or set new therapy.
Pressure high An occlusion occured in the system. The set
pressure level was exceeded. A bolus reduction is
automatically initiated by the pump.
Check if syringe is empty, kinks are in tubing and
tubing isn't damaged, IV patency and filter patency.
Increase occlusion pressure if necessary.
Due to varying syringe tolerances of syringes from
other manufacturers, a pressure alarm may occur
because of high syringe friction forces.
Syringe not correctly inserted” The wings of the syringe are not properly inserted.
Insert syringe according to describtion in Overview
Perfusor® Spaceas well as 1.1.
Syringe holder The syringe holder was opened during a running
infusion. Close syringe holder.
Battery cover removed The battery cover is not properly engaged on the
battery compartment.
When pushing on the battery cover listen for click”.
Drive blocked An external interference kept the drive unit from
advancing.
Basically prevent all external interferences.
Consider “Patient Safety“.
Chapter 6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Perfusor Space and is the answer not in the manual?

Braun Perfusor Space Specifications

General IconGeneral
Device TypeInfusion Pump
Accuracy±2%
Bolus RateAdjustable
Power Supply100-240 V AC, 50/60 Hz
DisplayLCD
AlarmsSystem error
Occlusion PressureAdjustable
Battery LifeApprox. 8 hours at 25 ml/h
CompatibilityB. Braun Space System
CertificationCE
Volume Preselection0.1 to 9, 999 ml