Errors and troubleshooting
Situation
No lens filter is attached.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Ambient temperature is not within
the allowed operating range
(10 40 °C or 50 104 °F).
Temperature taken is not within
typical human temperature range
(34 42.2 °C or 93.2 108 °F).
HI = too high
LO = too low
System error – self-check display flashes
continuously and will not be followed by
the ready beep and the ready symbol.
If error persists,
If error still persists,
Battery is low, but thermometer will still
operate correctly.
Battery is too low to take correct
temperature measurement.
Error Message
Solution
Attach new, clean lens filter.
Take care that the positioning of the
probe is correct and remains stable.
Replace lens filter and reposition.
Press Start button to begin a new
measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10
and 40 °C or 50 and 104 °F .
Make sure the probe tip and lens are
clean and a new, clean lens filter is
attached. Make sure the thermometer
is properly inserted. Then, take a new
temperature.
Wait 1 minute until the thermometer
turns off automatically, then turn
on again.
... reset the thermometer by re-moving
the batteries and putting them back in.
please contact the service centre.
Insert new batteries.
Insert new batteries.
Product specifications
Displayed temperature range: 34 – 42.2 °C (93.2 – 108 °F)
Operating ambient temperature range: 10 – 40 °C (50 – 104 °F)
Storage temperature range: -25 – 55 °C (-13 – 131 °F)
Operating and storage relative humidity: 10 – 95 %RH (non condensing)
Display resolution: 0.1 °C or °F
Accuracy for displayed temperature range Maximum Laboratory Error
35 °C 42 °C (95 °F 107.6 °F): ± 0.2 °C 0.4 °F)
outside this range: ± 0.3 °C 0.5 °F)
clinical repeatability: ± 0.14 °C 0.26 °F)
Battery life: 2 years / 350 measurements
Service life: 5 years
This thermometer is specified to operate at 1 atmospheric pressure or at altitudes with an
atmospheric pressure up to 1 atmospheric pressure (700-1060hPa).
See Instruction for use
Equipment with type
BF applied parts
10 °C
40 °C
Operating temperature
Keep it dry
Subject to change without notice.
This appliance conforms to the following standards:
Standard Reference Edition Title:
EN 12470-5: 2003 Clinical thermometers - Part 5: Performance of infra-red ear thermometers (with
maximum device).
EN 60601-1: 2006 Medical electrical equipment Part 1: General requirements for basic safety and
essential performance.
EN ISO 14971: 2012 Medical devices Application of risk management to medical devices.
EN ISO 10993-1: 2009 Biological evaluation of medical devices – Part 1:
Evaluation and Testing within a risk management process.
EN 60601-1-2: 2007 Medical electrical equipment part 1-2: General requirements for basic safety
and essential performance
Collateral standard: electromagnetic compatibility
Requirements and tests
EN 980: 2008 Symbols for use in labeling of medical devices.
EN 1041: 2008 Information supplied by the manufacturer of medical devices.
EN 60601-1-11: 2010 Medical electrical equipment -- Part 1-11: General requirements for basic
safety and essential performance -- Collateral standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare environment.
This product conforms to the provisions of the EC directive 93/42/EEC.
Medical electrical equipment needs special precautions regarding EMC. For detailed description of
EMC requirements please contact your authorized local service centre (See guarantee card).
Portable and mobile RF communications equipment can affect Medical electrical equipment.
Please do not dispose of the product in the household waste at the end of it useful life.
To protect the environment, dispose of empty batteries at appropriate collection sites
according to national or local regulations.