EasyManua.ls Logo

Brigon 500 NT - Technical Data

Default Icon
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BRIGON 500 NT Page 18
Technical data
Flue gas temperature
Range: -0°C .. + 600°C Type K, 0..400 °C
TÜV approved acc. to EN 500379-1 and -2
Resolution: 0,1 °C
Accuracy:
± 2 °C (up to 125 °C)
± 3 °C (125...250 °C)
± 4 °C (from 250 °C)
t0,98-time 50s
Ambient air temperature
Range: -0°C ... + 100°C Type K, 0...80°C
TÜV approved acc. to EN 500379-1 and -2
Resolution: 0,1 °C
Accuracy:
± 1 °C, t
0,98
-time 120s
CO
2
concentration
Range: 0...20,0 Vol% CO
2
TÜV approved acc. to EN 500379-1 and -2
Resolution: 0,1 Vol%
Accuracy:
± 0,2 Vol%, t
0,97
-time 50s
CO concentration
Range: 0...2000 ppm
TÜV approved acc. to EN 500379-1 and -3
short-time overstressing up to 4000 ppm possible
Resolution: 1 ppm
Accuracy:
± 20 ppm (up to 400 ppm)
± 5 % from reading (over 400 ppm)
Pressure and Draught
Range: -80 mbar ... + 80 mbar
TÜV approved acc. to EN 500379-1 and -2
Resolution: 0,01 mbar
Accuracy:
± 0,02 mbar (up to ± 5 mbar)
± 1 % from reading (over ± 5 mbar)
Additional Specifications
Ambient operating
range:
+ 5 ... + 40 °C
Transport and
storage temperature:
- 20 ... + 50 °C
Approval: TÜV By RgG 267 for O
2
, T
flue gas
and T
ambient air