Adding the product to Bronkhorst FlowSuite
When Bronkhorst FlowSuite is started, the software will
automatically search for connected instruments. All found
instruments are displayed in the left pane.
Use the refresh option to check for newly connected products.
If the product is not found, it may not be set to serial RS-232.
Follow the instructions on the next page to set the product
to serial RS-232.
Commissioning
Apply fluid pressure gradually to prevent pressure shocks [1].
Warm up the product for at least 30 minutes before starting
measurement [2].
Adjusting Zero Point
To ensure correct measurements, adjust Zero Point in the following
situations: before first use and after remounting [3].
To start adjusting Zero Point:
Open Bronkhorst FlowSuite on the PC.
Select the product in the left pane.
Go to the Actions menu.
Press the button Auto Zero.
Follow the on-screen instructions.
The product is ready for use!