3 .
2.1.
Measuring Procedures
Input
Signal
Measuring Ampl
if
ie
r 2606,
2607, 2608
Band
Pa
ss
Filter 1617
770356
Fig. 3,2.
Basic
instrumentation
arrangement
1. Connect
the
instrumentation
as
shown
in
Fig .3 .2 .
2. Perform
the
preliminary
adjustments
for
the
Measuring
Amplifier
as described in the
relevant
Instruction
Manual
.
3. Set
the
Measuring
Amplifier
controls
as
follows
:
INPUT
"Direct"
when
the
B & K coaxial
input
is used, or
"
Preamplifier
Input"
when
the
7 -pin B & K socket
is used
METER
FUNCTION
As
appropriate:
for
further
information
see the
Measuring
Amplifier
Instruction
Manual
GAIN CONTROL
"Calibrated"
INPUT ATIENUATOR
"300
V"
OUTPUT
ATIENUATOR
"x
1"
FILTERS "
External",
plus
any
weighted
network
required;
if
only
audio
frequencies
are
to
be measured,
the
high pass
22.4
Hz
filter
and the
low
pass
22.4
kHz
fil!er
(not on
2608)
can be used to
limit
the
frequency
range
of
the
Filter
output
signal.
REMOTE/
MANUAL
Averaging
.
Time
Switch
(rear panel
of
2607
only)
"Manual"
POWER
"On"
4 . Perform
the
preliminary
adjustments
for
the
Band Pass Filter as described in section
3.1.
5. Set
the
Band Pass Filter controls
as
follows:
SELECTIVITY
RANGE
AVERAGING CONTROL
FILTER
CONTROL MODE
Bandwidth
(and stepping) as required
As
appropriate
Switch
positions
immaterial
Left-hand
switch
"Manual"
Right-
hand
switch
in
neutral
position