18
Storage Temperature: -20
o
C to 60
o
C, < 80% R.H. (with battery removed)
Altitude: Operating below 2000m
Temperature Coefficient: nominal 0.15 x (specified accuracy)/
o
C @(0
o
C ~ 18
o
C or 28
o
C ~
45
o
C), or otherwise specified
Sensing:
TBM829 & TBM525: AC+DC True RMS
TBM812 & TBM812XEX: AC True RMS
TBM811 & TBM811XEX: Average Sensing
Safety: Double insulation per IEC61010-1 2nd Ed., EN61010-1 2nd Ed., UL61010-1 2nd
Ed. & CAN/CSA C22.2 No. 61010.1-0.92 to Category IV 1000Vac & Vdc.
Transient protection: 12kV (1.2/50s surge)
TBM811, TBM812 & TBM829 Terminals (to COM) Measurement Category:
V : Category IV 1000Vac & Vdc
mAA : Category IV 600Vac and 300Vdc
A : Category IV 600Vac and 300Vdc
TBM811XEX, TBM812XEX & TBM525 Terminals (to COM) Measurement Category:
V / A / mAA : Category IV 1000Vac & Vdc
Overload Protections:
TBM829, TBM812 & TBM811:
A & mA: 1A/600Vac, IR 10kA or better, F fuse
A: 10A/600Vac, IR 100kA or better, F fuse
V: 1050Vrms, 1450Vpeak
mV, , & Others: 600Vdc and Vac rms
TBM812XEX & TBM811XEX:
A & mA: 0.44A/1000Vac & Vdc, IR 10kA or better, F fuse
A: 11A/1000Vac & Vdc, IR 20kA or better, F fuse
V: 1050Vrms, 1450Vpeak
mV, , & Others: 600Vdc and Vac rms
TBM525:
A & mA: 0.44A/1000Vac & Vdc, IR 10kA or better, F fuse
A: 11A/1000Vac & Vdc, IR 20kA or better, F fuse
V, mV, , & Others: 1050Vrms, 1450Vpeak
E.M.C. : Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-3-3, EN61000-4-2,
EN61000-4-3, EN61000-4-4, , EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges:
Total Accuracy = Specified Accuracy + 100 digits
Performance above 3V/m is not specified
Power Supply: Single 9V battery; NEDA1604G, JIS006P IEC6F22, NEDA1604A, JIS6AM6