BÜCHI Labortechnik AG Help with faults | 8
Operation Manual Inert Loop B-295 SE 33/38
u Install the grounding cable.
u Install the front cover
8.3 Error messages
NOTICE
Too frequent switching on the instrument
Too frequent switching on the instrument can cause an instrument damage.
u Wait 10 minutes before restarting the instrument.
Fault code Fault Remedy
E01 Temperature sen-
sor defective
u Switch off the instrument.
u Wait until the instrument is in ambient temper-
ature.
u Remove dust and foreign objects from the
ventilation slots using compressed air or a
vacuum cleaner.
u Switch on the instrument.
u Contact BUCHI Customer Service.
E02 Temperature error
E04 Compressor pres-
sure fault
u Switch off the instrument.
u Wait until the instrument is in ambient temper-
ature.
u Switch on the instrument.
u Contact BUCHI Customer Service.
E05 Data error u Switch off the instrument.
u Switch on the instrument.
u Contact BUCHI Customer Service.
E06 Electronic circuitry
overheated
u Switch off the instrument.
u Wait until the instrument is in ambient temper-
ature.
u Remove dust and foreign objects from the
ventilation slots using compressed air or a
vacuum cleaner.
u Switch on the instrument.
u Contact BUCHI Customer Service.