11
Digestion Unit K-435 5 Operation
5.4 Conducting Kjeldahl Digestion
Turn on the apparatus a and preheat it at Setting 10 for
approx. 5 minutes.
Wear protective goggles!
Place the digestion glasses in the digestion glass holder b
on the rack c.
Add the sample and the reagents to the digestion glasses
d.
Close the unused connecting pieces with plug caps
(Art. No. 17396).
Never operate the apparatus with
empty digestion glasses!
Place the suction module e with the mounted seals over
the digestion glass holder..
Turn on scrubber or water jet pump before any gases form.
5.5 Conducting Kjeldahl Digestion in
Remote Control Operation
Connect the Büchi Control Unit B-436 to the Digestion
Unit and turn on the apparatus.
Enter the parameters on the B-436.
Set the regulator on the Digestion Unit at Setting 0.
Place the digestion glasses in the digestion glass holder
on the rack.
Add the samples and the reagents to the digestion glasses.
Close the unused connecting pieces with plug caps
(Art. No. 17396).
Never operate the apparatus with
empty digestion glasses!
Place the suction module e with the mounted seals over
the digestion glass holder.
Begin digestion by pressing the START button on the
Control Unit B-436 (Scrubber B-414 starts automatically;
manually start the water jet pump before any gases form).
b
c
d
a
Figure 10: Operation of K-435
e