Digestion Unit K-4355 Operation
5.6 Malfunctions
Malfunction Possible Cause Correction
Heating elements 1. Apparatus is not connected to - Connect the apparatus to the power supply
do not heat power supply
2. Apparatus is turned off - Turn off apparatus
3. Heat regulator is at setting „0“ - Set to setting "10"
4. Heat regulator is defective - Replace regulator
5. Heating elements are defective - Replace heating elements
Light does not Light is defective Replace light
come on (H 1)
Table 4: Malfunctions