6 Operation
44
Melting Point M-565 Operation Manual, Version F
User calibration:
By choosing the user cali-
bration, 4 freely selectable
substances can be chosen for
calibration.
Choose the gradient for the
calibration with the rotary
knob.
Insert the name of the
substance with starting the
lowest melting point used for
the calibration.
Set the melting point value for
the substance using the rotary
knob.
Enter the Lot number for the
substance used.
Enter the correct reference
temperature for the substance
according to the data sheet.
This value is used for the
actual calibration. Measure-
ment as described for BUCHI
calibration.
The calibration date is now
saved and indicated on the
idle screen..