5 | Installation
Büchi Labortechnik AG
26/102
Operation Manual Pure Chromatography Instruments
Precondition:
R Make sure that the instrument is not
connected to the power supply.
u Attach all solvent lines to the
instrument. Connections see
Chapter3.2.4 "Connections on gradient
valve", page17
u Put the other end of the solvent line
into the solvent bottle.
u Assign the solvent to the solvent lines.
See Chapter5.7 "Assigning solvent to
solvent lines", page26
u Install the solvent level sensor. See
Chapter5.6 "Installing the solvent level
sensor", page26
5.6 Installing the solvent level sensor
Precondition:
R Make sure that the instrument is not
connected to the power supply.
u Attach all solvent level sensors to the
instrument. Connections see
Chapter3.2.4 "Connections on gradient
valve", page17
u Calibrate the solvent level sensors. See
Chapter5.8 "Calibrating the solvent
level sensor", page27
5.7 Assigning solvent to solvent lines
Navigation path
Tools Solvent Loading
Precondition:
R The solvent bottle is connected to the instrument. See Chapter5.5 "Establishing
solvent connection", page25