5.12 Calibrating the waste level sensor
Navigation path
➔ Tools ➔ Calibration and Defaults
Precondition:
R The waste level sensor is not covered by liquid waste.
u Navigate to the
Calibration
dialog according to the navigation path.
ð The display shows the dialog box
Calibration
.
u Tap [Zero] for [Waste].
u Tap the button [Close].
ð The dialog box closes.