7 Maintenance and repairs
46 K-425 / K-436 Operation Manual, Version B
7.5.2 500 ml sample tube support spring
The 500 ml tube rack is equipped with a tube holder containing five sample tube support springs.
Each flat spring must be freely rotatable by a small amount. If swivelling is not possible, exchange the
rake!
Note
Rinse the tube holder with water after use to avoid corrosion.
Store it dry and clean after use.
7.6 Protection cover at the front
The protection cover at the front is equipped with double-faced adhesive tape. If it is strongly contami-
nated or damaged, replace it with a new one.