Atria Service Manual
iii
Warnings & Cautions
Warnings
WARNING:
There can be high voltage near the fuse, AC inlet and
transformer. Always unplug the electrocardiograph before taking it
apart.
Cautions
CAUTION:
Always turn the electrocardiograph off and disconnect
the power cord before cleaning. Do not pour liquids (such as alcohol or
other cleaners) on the unit. This will cause severe electrical damage!
CAUTION:
Setting the printhead resistance to any value other than
the value written on the printhead will shorten the life expectancy of
the printhead and may void your warranty.
CAUTION:
Several of the electrocardiograph’s components are
extremely sensitive to static electricity. To minimize the risk of damage,
use a grounding strap and work on an antistatic surface.
CAUTION:
The LCD is very sensitive to static electricity. Use an
antistatic work surface.
CAUTION:
The queue sensor is very sensitive to static electricity. Use
an antistatic work surface.
CAUTION:
The printhead is very sensitive to static electricity. Use an
antistatic work surface.
CAUTION:
The optional I/O board is very sensitive to static
electricity. Use an antistatic work surface.
CAUTION:
The front end digital board is very sensitive to static
electricity. Use an antistatic work surface.
CAUTION:
The power supply board is very sensitive to static
electricity. Use an antistatic work surface.