What to do if Burdick ELI 280 says 'A link to the access point was established, but the link to the destination failed'?
- CCatherine HaydenSep 23, 2025
Ensure the IP address is correct. If problem persists, contact Mortara Technical Service.
What to do if Burdick ELI 280 says 'A link to the access point was established, but the link to the destination failed'?
Ensure the IP address is correct. If problem persists, contact Mortara Technical Service.
What to do if unable to transmit ECG on Burdick ELI 280?
If you are unable to transmit an ECG with your Burdick Medical Equipment, verify the phone number is correct and ensure the modem is online.
Why is Burdick ELI 280 unable to transmit ECG?
If you are unable to transmit an ECG with your Burdick Medical Equipment, check the phone line and ensure the site number is valid. Then, try again.
How to fix Burdick ELI 280 Medical Equipment when the device is not responding?
If your Burdick Medical Equipment device is not responding, press and hold the On/Off button for 10 seconds. Note that re-entry of date and time will be required after performing this action.
What to do if Burdick ELI 280 shows a lead fail error?
If the Burdick Medical Equipment indicates a 'Lead fail,' correct the faulty lead.
Why does Burdick Medical Equipment display 'Received orders failed'?
If received orders failed on your Burdick Medical Equipment, it may be because the previous Custom ID is not compatible with the current Custom ID, or there is no Custom ID.
What to do if DICOM order storage failed on Burdick ELI 280?
If DICOM order storage failed on your Burdick Medical Equipment, the directory is full; mark records for deletion or delete records.
How to fix 'Unable to connect with LAN with crossover cable' error on Burdick ELI 280?
If you are unable to connect with LAN with a crossover cable on your Burdick Medical Equipment, implement a hub instead of a crossover cable.
What to do if the AC power cord of Burdick ELI 280 is not connected or damaged?
If the AC power cord is not connected to a grounded electrical outlet or is damaged, ensure that the AC power cord is not damaged and is firmly connected to the AC power connector on the rear of the electrocardiograph. Ensure that the electrocardiograph is plugged into a grounded electrical outlet. If AC power is being used and the AC power switch is set to the On position, but the AC power on indicator light does not illuminate and the display is still dark, contact Mortara Technical Support.
What to do for poor LL electrode or left leg tremor on Burdick ELI 280 Medical Equipment?
If your Burdick Medical Equipment is showing a poor LL electrode or left leg tremor, check the patient prep and re-prep if necessary using a new electrode.
Details on Mortara Instrument's responsibility for product safety and performance.
Information regarding document updates, warranties, and disclaimers.
Contact information for Mortara Instrument's main headquarters.
Contact details for technical assistance and support services.
Details on the warranty coverage for Mortara products and limitations.
Critical safety warnings regarding device operation and potential hazards.
Precautions to prevent device damage or reduced usability.
Information about wireless data transmission and RF safety guidelines.
Explanation of various symbols and markings used on the equipment.
General precautions for handling and cleaning the device.
Procedures for cleaning and disinfecting the device and its accessories.
Manufacturer's declaration regarding electromagnetic emissions compliance.
Guidelines for safe distances from RF equipment to prevent interference.
Overview of the manual's objective and the information it provides.
Description of the ELI 280 electrocardiograph and its capabilities.
Explanation of the ELI 280's display screen and its parameters.
Steps for initial setup and configuration of the ELI 280.
Instructions for correctly loading thermal paper into the printer.
Details on operating the device using AC power and battery.
Steps to prepare the patient before acquiring an ECG.
Guidance on proper electrode placement for accurate ECG readings.
Procedures for entering patient demographic information.
Options and procedures for transmitting ECG data.
Instructions for connecting the device to a wired network.
Steps for configuring and connecting to a wireless network.
Overview of the ECG directory for storing and managing records.
Information on managing ECG orders and the worklist.
Methods for searching ECG records by patient name or ID.
Overview of system settings and access to configuration menus.
Guidance on setting and managing device administrator and technician passwords.
Settings related to ECG acquisition, display, and filtering parameters.
Lists common LCD messages, their problems, and corrective actions.
Troubleshooting guide for ECG signal artifacts and lead misplacements.
Recommendations for maintaining battery life and proper charging practices.