REF 9515-181-51-ENG Rev G
ELI 280
RESTING ELECTROCARDIOGRAPH
USER MANUAL
Manufactured by Mortara Instrument, Inc., Milwaukee, Wisconsin U.S.A.
CAUTION: Federal law restricts this device to sale by or on the order of a physician.

Table of Contents

Questions and Answers

  • C
    Candace GalvanSep 8, 2025
    What to do if Burdick ELI 280 Medical Equipment is unable to print, paper queue mark not detected?
    • M
      Michael AguilarSep 8, 2025
      If your Burdick Medical Equipment is unable to print because the paper queue mark was not detected as expected, add paper. Then, manually advance the page evenly past the closure point of the writer, close the writer cover, and press STOP.
  • M
    Mason DaviesSep 7, 2025
    How to turn on Burdick Medical Equipment when it is in Standby Mode?
    • H
      hmartinSep 7, 2025
      If your Burdick Medical Equipment is in Standby Mode, press the On/Standby button to return to active use.
  • M
    Margaret DickersonAug 22, 2025
    How to fix Burdick Medical Equipment when unable to acquire ECG or print?
    • S
      Steven HawkinsAug 22, 2025
      If you are unable to acquire an ECG or print with your Burdick Medical Equipment, charge the battery using AC power.
  • V
    Vincent LopezSep 4, 2025
    How to resolve poor LA electrode or left arm tremor on Burdick ELI 280?
    • D
      Donald DavisSep 4, 2025
      If your Burdick Medical Equipment indicates a poor LA electrode or left arm tremor, check the patient prep and re-prep if necessary with a new electrode.
  • S
    Susan GutierrezSep 3, 2025
    What to do for poor LL electrode or left leg tremor on Burdick ELI 280 Medical Equipment?
    • R
      Ryan OlsonSep 3, 2025
      If your Burdick Medical Equipment is showing a poor LL electrode or left leg tremor, check the patient prep and re-prep if necessary using a new electrode.
  • C
    christophernealSep 2, 2025
    What causes poor RA electrode or right arm tremor in Burdick ELI 280?
    • R
      Rachel HernandezSep 2, 2025
      If you are getting a reading of poor RA electrode or right arm tremor on your Burdick Medical Equipment, check the patient prep and re-prep if necessary with a new electrode.
  • L
    Lauren DavisAug 31, 2025
    How to eliminate high frequency noise in Burdick ELI 280 Medical Equipment?
    • D
      Destiny JonesAug 31, 2025
      If you are experiencing high frequency noise with your Burdick Medical Equipment, adjust the low pass filter setting; check proximity to power cables; and check the AC filter setting (50 Hz or 60 Hz).
  • S
    stephanie41Aug 30, 2025
    Why is Burdick Medical Equipment unable to transmit or receive ECGs?
    • S
      samantha70Aug 30, 2025
      If you are unable to transmit or receive ECGs on your Burdick Medical Equipment, check for the correct baud rate, phone number, and cable connections or site number.
  • J
    Jasmine BrownAug 29, 2025
    What to do if Burdick ELI 280 Medical Equipment says 'No available memory. ECG data too noisy to store'?
    • W
      Wendy BarnettAug 29, 2025
      If your Burdick Medical Equipment displays 'No available memory' because the ECG data is too noisy to store, press stop to continue. Transmit or mark records for deletion in the directory. Correct the noise issue and attempt acquisition/storage again.
  • S
    Shawn PhamAug 27, 2025
    What to do if Burdick ELI 280 shows a lead fail error?
    • S
      scottwilliamsonAug 27, 2025
      If the Burdick Medical Equipment indicates a 'Lead fail,' correct the faulty lead.

Summary

Notices

Manufacturers Responsibility

Details on Mortara Instrument's responsibility for product safety and performance.

Other Important Information

Information regarding document updates, warranties, and disclaimers.

Technical Support and Service

Headquarters

Contact information for Mortara Instrument's main headquarters.

Service;Technical Support Group

Contact details for technical assistance and support services.

Warranty Information

Your Mortara Warranty

Details on the warranty coverage for Mortara products and limitations.

User Safety Information

Warning(s)

Critical safety warnings regarding device operation and potential hazards.

Caution(s)

Precautions to prevent device damage or reduced usability.

Wireless Data Transmission

Information about wireless data transmission and RF safety guidelines.

Equipment Symbols and Markings

Symbol Delineation

Explanation of various symbols and markings used on the equipment.

General Care

Precautions

General precautions for handling and cleaning the device.

Cleaning and Disinfecting

Procedures for cleaning and disinfecting the device and its accessories.

Electromagnetic Compatibility (EMC)

Guidance and Manufacturers Declaration: Electromagnetic Emissions

Manufacturer's declaration regarding electromagnetic emissions compliance.

Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the Equipment

Guidelines for safe distances from RF equipment to prevent interference.

Introduction

Manual Purpose

Overview of the manual's objective and the information it provides.

System Description

Description of the ELI 280 electrocardiograph and its capabilities.

Display Overview

Explanation of the ELI 280's display screen and its parameters.

Equipment Preparation

Initial Startup

Steps for initial setup and configuration of the ELI 280.

Loading Paper

Instructions for correctly loading thermal paper into the printer.

Powering the ELI 280

Details on operating the device using AC power and battery.

Record an ECG

Patient Preparation

Steps to prepare the patient before acquiring an ECG.

Patient Hookup

Guidance on proper electrode placement for accurate ECG readings.

Patient Demographic Entry

Procedures for entering patient demographic information.

Connectivity and ECG Transmission

ECG Transmission

Options and procedures for transmitting ECG data.

Local Area Network (LAN) Connection and Setup

Instructions for connecting the device to a wired network.

Wireless Local Area Network (WLAN) Connection and Setup

Steps for configuring and connecting to a wireless network.

ECG Directory

ECG Directory

Overview of the ECG directory for storing and managing records.

ECG Orders

Information on managing ECG orders and the worklist.

Searching ECG Records

Methods for searching ECG records by patient name or ID.

System Settings

Menu Commands and Utilities

Overview of system settings and access to configuration menus.

Device Passwords

Guidance on setting and managing device administrator and technician passwords.

Configuration Settings: ECG

Settings related to ECG acquisition, display, and filtering parameters.

Maintenance and Troubleshooting

System Troubleshooting Chart

Lists common LCD messages, their problems, and corrective actions.

ECG Troubleshooting Chart

Troubleshooting guide for ECG signal artifacts and lead misplacements.

Battery Maintenance

Recommendations for maintaining battery life and proper charging practices.

Burdick ELI 280 Specifications