SYSTEM SETTINGS
71
Communication Protocol
This control allows the IT professional to set the communication protocol to UNIPRO32, DICOM32, or
DICOM32ext.
NOTE: This setting must be entered under the direction of the IT professional at the facility where the
device is installed.
Caps Lock
All character entry is translated to uppercase.
Transmitted Edit Disable
When set to Yes, the ELI 280 will not allow an ECG’s demographics to be changed after it has been transmitted.
Barcode Date Format
Distinguishing between months and days in formatted date strings is not always possible. When a barcode scanner
is used to read dates, this setting specifies if dates are in MM/DD (month/day) or DD.MM (day.month) format.
Cart Number
This control allows the user to assign an ELI 280 with a cart number from 0 to 65535 in order to identify which
device acquired or transmitted a particular ECG.
Site Number
This control allows the user to specify the location of the ELI 280 using a site number. Site numbers designate the
department, nursing unit, hospital, clinic, or institution for ECG records stored in an E-Scribe or approved third-
party cardiology management system and must be defined for transmitting and retrieving ECGs from that system.
Up to four digits (0–4095) are supported to assign a site number. Up to three sites can be defined when multi
protocol is enabled.
Site Name
This control allows the user to name the department, nursing unit, clinic, hospital or office where the ELI 280 is
located. The site name prints at the bottom, left edge of the ECG printout. Use up to 30 alphanumeric characters for
the name. Up to three sites can be defined when multi protocol is enabled.
Output Date Format
This control allows the user to define the date of birth format output to an E-Scribe in a custom ID.
Mandatory Fields
These controls specify if the patient’s ID, patient’s last name, and/or the technician’s ID must be valued before an
ECG can be acquired or transmitted.
Sync Mode
This control allows the user to set the transmission option when synchronizing with the electronic medical record
system. Synchronization can be set to None, Transmit, Transmit+Orders, or Transmit+Orders+Date/Time.
NOTE: This device supports automatic time synchronization with a remote server. Inaccurate
time/date settings can result in ECGs marked with inaccurate time stamps. Confirm the accuracy
of the synchronized time prior to acquiring ECGs