Sensor
excitation
[V]
Output
[mV/V]
Calculated
bridge
voltage
[mV]
Output
(reference)
[V]
Calculated
gain
Stage 1 Set
gain
(decimal)
Stage 2
Set gain
(0.25…63)
Stage 3
Set gain
(2/4/6/8)
10
0.1
1
10
10.000.00
100
50
2
10
0.2
2
10
5.000.00
100
25
2
10
0.5
5
10
2.000.00
100
10
2
10
1
10
10
1.000.00
100
5
2
10
1.25
12.5
10
800.00
100
4
2
10
1.5
15
10
666.67
10
33.25
2
10
1.75
17.5
10
571.43
10
28.5
2
10
2
20
10
500.00
10
25
2
10
2.5
25
10
400.00
10
20
2
10
3
30
10
333.33
10
16.75
2
Table 6: Gain setting for strain-gauge full-bridge sensors