ACCESSORIES / SPARE PARTS
SEPARATLY AVAILABLE SPARE PARTS
DRUG CHAMBER REF SP 0279
Soft Touch K+ MASK (L) ADULTS REF SP 0281
Soft Touch K+ MASK (S) CHILD - REF SP 0282
Soft Touch K+ MASKS CONNECTOR REF SP 0283
MOUTHPIECE REF SP 0263
K+ NOSEPIECE REF SP 0284
AC/DC POWER SUPPLY - REF SP 0262
AC/DC POWER SUPPLY REF SP 0262/01
EVO Black Carrying Case for KIWI Plus - REF SP 0280
DRUG CHAMBER: For each single patient, we recommend replacing the drug chamber ever 12 months or 350
treatments at the most.. In addition, the drug chamber must be replaced in the case of cracks or if the holes in the
Vibrating Mesh inside it are partially or totally obstructed by medication, dust, limestone deposits, etc..
In the presence of diseases with risk of infection and microbial contamination, we recommend a personal use of
the accessories and drug chamber and compliance with the sterilisation procedures described in this manual
(always consult your medical specialist).
Masks and mouthpieces must be replaced if they show visible signs of decay.
Minimum Expected service life: 2 Years for main unit and power supply MAX 180 operating hours for drug
chamber in accordance with the standard test and operation conditions.
Shelf life: max 5 years from the date of manufacture but max 2 YEARS without application of a full recharging of
internal battery
POSSIBLE PROBLEMS AND SOLUTIONS
Defect
Cause
Remedy
1. Very low or no
nebulisation (with device
on)
The holes in the vibrating
mesh, located inside the
drug chamber, are clogged
The backlight of Power
Button flashed quickly
(the interval is less than
0.3 second), and the
batteries are low/worn out.
Clean the drug chamber as instructed in the
manual. If the problem persists even after
cleaning, replace the drug chamber with a
new one
Recharge the internal battery (full recharging
cycle)
2. No nebulisation with
device on (power button
backlight is on)
The electrodes in the drug
chamber or in the main
unit are dirty and/or
stained
Medication or water is
collected on the Electrodes
of the Main Unit.
Clean the contacts by removing the cause of
the poor electric contact.
Remove the medication or water.
Replace the Medicine Cup with a new one.
Only use original
accessories/spare
parts provided and
indicated by the

Related product manuals