AEROSOL OUTPUT (EN 13544-1)
AEROSOL OUTPUT RATE
(EN 13544-1)
0.30 ml/min (2ml NaCl 0.9%)
INSULATION CLASS
Class II (if used with HNEL050100WE or SC06W-0501000V
power supply)
Internally Powered Equipment (if used with internal batteries)
Approx.150 g (without accessories)
48 (W) x 127 (H) x 58 (D) mm
Room temperature: 10 ÷ 40 °C
Relative humidity: 30 ÷ 85 %RH (non-condensing)
Atmospheric pressure: 700 - 1060 hPa
AND OF TRANSPORT
Room temperature: -20 ÷ 50 °C
Relative humidity: 30 ÷ 80 %RH (non-condensing)
Atmospheric pressure: 795 - 1013 hPa
* With new battery, intact and with 100% of charge available
** MMAD = Mass Median Aerodynamic Diameter (measured with Marple Personal Cascade Impacter)
*** GSD = Geometric Standard Deviation
**** The parameters shown are used in production during 100% verification of the devices as acceptance limits and
therefore represent the minimum threshold value to ensure compliance with the parameters measured in accordance
with EN 13544-1