Stroller/Sprint
10
Warning: Do not operate Stroller/Sprint while
attached to the stationary unit.
Warning: Should the leakage be excessive to the
point that a stream of liquid is present, leave the area
and call your health care provider immediately.
Caution: If the Stroller/Sprint is being refilled
immediately after a period of use, fill it, and then
wait 10 minutes after filling with the FCV set at zero
for the pressure to stabilize. Then operate as
normal.
Warning: Liquid or cold gaseous oxygen can frostbite
the skin. Oxygen will result in increased fire hazard.
Portable
Release
Button
(push-on
style only)
Closing the valve at the first sign of liquid discharge
does not fill the Stroller/Sprint as completely, but saves
liquid oxygen.
9. For the push-on style connectors, push the release
button down until the units separate. For all other
connectors rotate the unit until it separates from the
stationary unit.
Caution: Check liquid level gauge only after the vent
valve is closed.
Caution: If the Stroller/Sprint does not separate easily,
do not use force. The units may be frozen together.
Wait until the units warm up then they’ll separate
easily. Do not touch any frosted parts. Should there be
any liquid leakage from the stationary unit or
Stroller/Sprint after separation, reattach the
Stroller/Sprint immediately. Again separate the units. If
the problem persists, leave units coupled and call your
health care provider.