To retain a numerical calibration setting and advance to the next menu without pressing ENTER/STEP
repeatedly, press and hold ACCESS for 3 seconds and release. This shortcut applies only to numerical
fields. You must press ENTER/STEP to select all other entries.
Data Entry Tip
Installing the Pulse Output Circuit Assembly
A 6-pin terminal strip and two mounting holes in the pulse output circuit assembly allow for easy installation.
To install the pulse output option, perform the following steps:
1.
Confirm that the W1 jumper is set in the ā1ā position. This setting allows you to configure the pulse out-
put scale using the keypad on the front of the enclosure.
2.
Route the field wiring through the supplied cord connector or the enclosure.
3.
Connect the wiring in accordance with Figure B2.
Choosing a Scale Factor
The scale factor that you enter into the MC-II sets the volume increment that will cause a pulse output to oc-
cur. The scale factors are:
0.001 One pulse per 0.001 volume increment
0.01 One pulse per 0.01 volume increment
0.1 One pulse per 0.1 volume increment
1.0 One pulse per 1 volume increment
10.0 One pulse per 10 volume increments
100.0 One pulse per 100 volume increments
Configuring the Pulse Output Option
To use the pulse output option, you must enable the pulse output feature and enter a pulse output scale factor
using the keypad menu, as described in the following steps:
1. Press and hold ACCESS for 3 seconds, then release, to enter the Calibrate mode (see Data Entry tip
below).
2. Proceed through the various menu displays, pressing ENTER/STEP to advance to the next menu, until
the prompt PULSE appears in the top display. The default setting is OFF.
3. Press INCREMENT to select the appropriate scale factor (0.001, 0.01, 0.1, 1.0, 10.0, 100.0). The scale
factor is the volume increment that will cause a pulse output to occur.
4. Press ENTER/STEP to accept the selection. The CodE menu will appear in the top display.
5. Press ENTER/STEP to accept the existing CodE setting, or change the password security setting and
then press ENTER/STEP. The SEt.tot menu (for presetting a total) will appear in the top display.
6. Press ENTER/STEP to bypass this menu selection or press INCREMENT to enable the preset total op-
tion. (See step 7 of the calibration procedure on page 16 for details on presetting a total.)
The prompt SAVing will appear in the lower display while the calibration settings are saved to nonvola-
tile memory. The MC-II will automatically return to the Run mode.