MODEL 199
DIFFERENTIAL PRESSURE UNIT (DPU)
Standard and NACE Service
User Manual
Manual No. 9A-10030, Rev. 01
January 2008
CONTENTS
Section 1 - Introduction .................................................................. Page 3
Section 2 - Theory of Operation .................................................... Page 5
Section 3 - Installation/Operation ................................................ Page 9
Section 4 - Maintenance, Adjustment, & Calibration ................. Page 13
4-1. Tools Required for Maintenance and Calibration ............. Page 13
4-2. DPU Inspection and Cleaning............................................. Page 13
4-3. Calibration Setup .................................................................. Page 18
4-4. Torque Tube Rotation Check (Replacement Units) ......... Page 18
4-5. Range Change ....................................................................... Page 19
4-6. Setting Bellows Travel .......................................................... Page 22
4-7. Bellows Unit Assembly (BUA) Replacement .................... Page 22
4-8. Attaching Drive Arm to Torque Tube ................................. Page 24
4-9. Drive Arm Tightness Test .................................................... Page 24
4-10. Adjusting Pulsation Dampener ......................................... Page 25
4-11. Troubleshooting .................................................................. Page 25
Section 5 - Parts Drawing/List ....................................................... Page 27
Section 6 - Outline Dimension Drawings ..................................... Page 30
(This manual is for the DPU only.) Refer to the appropriate (separate)
instrument manual for information on the instrument being actuated.

Questions and Answers

  • T
    Teresa AllenAug 12, 2025
    Why is my Cameron Barton 199 Measuring Instruments showing a high indication?
    • H
      Hannah SimpsonAug 12, 2025
      If your Cameron Measuring Instruments are showing a high indication, it could be due to a partially restricted or too small orifice; clean out or replace it. Another cause could be a leak in the LP side piping, so repair any leaks. Gas trapped in the LP housing (liquid service) or liquid trapped in the HP housing (gas service) can also cause this; vent the housing. If the LP housing gasket is leaking, replace it. A broken range spring or a tampered DPU might also be the cause, in which case you should return the BUA for repair. Repair or replace any loose links or movements, or calibrate the instrument if it's out of calibration.
  • A
    Amy WebbAug 22, 2025
    What to do if Cameron Barton 199 show erratic indication?
    • K
      Kathleen HughesAug 22, 2025
      If your Cameron Measuring Instruments are showing an erratic indication, it may be due to flow pulsation. Installing a dampening device upstream of the DPU run may help. Liquid trapped in gas piping or a gas bubble in liquid piping could also be the problem; remove it according to the Startup Procedure in Section 3. If the vapor generator is incorrectly installed, repipe it. If the reference leg is gassy or the liquid is vaporizing, refer to the piping instructions. Obstructions to the bellows travel can cause erratic readings, so clean the bellows. If gas is trapped in the DPU HP or LP housing, remove it according to the Startup Procedure in Section 3. Adjust or clean any dragging or dirty linkages, and adjust the pointer if it's dragging on the scale plate.
  • Q
    qjacksonNov 2, 2025
    What to do if Cameron Barton 199 Measuring Instruments show low or no indication?
    • M
      Michael DukeNov 2, 2025
      If your Cameron Measuring Instruments show low or no indication, it could be due to several reasons. The orifice might be installed backward or be oversized, in which case you should replace it. There might be a blockage upstream from the run, so clean out the run or open the valve. Another cause could be loss of liquid in the reference leg, so refill it. Density changes in the process media or reference leg can also cause this issue; refill the reference leg with the same density liquid as the process media. Also, check for plugged pressure tap holes and clean out the piping, or ensure the bypass valve is closed and not leaking. Clear any trapped liquids or gases by venting the piping, and make sure block or shut-off valves are open. Repair any piping leaks on the HP side. If the flapper ...

Summary

General Safety Warnings and Precautions

Theory of Operation

Installation and Operation

DPU Mounting Procedures

Guidance on flush/panel mounting and pipe mounting the DPU.

Piping Standard Practices

Recommended practices for piping connections to the DPU.

General Startup Practices

Key practices for safely starting up the instrument.

Maintenance, Adjustment, and Calibration

Pressure Check Procedure

Procedure to check for internal pressure before disassembly.

DPU Range Change Procedure

Procedure for changing the range of the Model 199 DPU.

Range Change for 2-1;8-inch Bellows

Steps to change the range for 2-1/8-inch diameter bellows units.

Range Change for 3-3;4-inch Bellows (No Kickoff Spring)

Procedure for changing range on 3-3/4-inch bellows without kickoff spring.

Range Change for 3-3;4-inch Bellows (With Kickoff Spring)

Procedure for changing range on 3-3/4-inch bellows with kickoff spring.

Bellows Unit Assembly (BUA) Replacement

Procedure for replacing the Bellows Unit Assembly (BUA).

Troubleshooting Guide

Guide to identify and correct common DPU problems.

Parts Drawing and List

Outline Dimension Drawings

Cameron Barton 199 Specifications